Combined experimental and Monte Carlo verification of brachytherapy plans for vaginal applicators
- 1 December 1998
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 43 (12) , 3495-3507
- https://doi.org/10.1088/0031-9155/43/12/008
Abstract
Dose rates in a phantom around a shielded and an unshielded vaginal applicator containing Selectron low-dose-rate 137Cs sources were determined by experiment and Monte Carlo simulation. Measurements were performed with thermoluminescent dosimeters in a white polystyrene phantom using an experimental protocol geared for precision. Calculations for the same set-up were done using a version of the EGS4 Monte Carlo code system modified for brachytherapy applications into which a new combinatorial geometry package developed by Bielajew was recently incorporated. Measured dose rates agree with Monte Carlo estimates to within 5% (1 SD) for the unshielded applicator, while highlighting some experimental uncertainties for the shielded applicator. Monte Carlo calculations were also done to determine a value for the effective transmission of the shield required for clinical treatment planning, and to estimate the dose rate in water at points in axial and sagittal planes transecting the shielded applicator. Comparison with dose rates generated by the planning system indicates that agreement is better than 5% (1 SD) at most positions. The precision thermoluminescent dosimetry protocol and modified Monte Carlo code are effective complementary tools for brachytherapy applicator dosimetry.Keywords
This publication has 24 references indexed in Scilit:
- Photon mass attenuation and energy-absorption coefficientsPublished by Elsevier ,2002
- Two-dimensional scatter integration method for brachytherapy dose calculations in 3D geometryPhysics in Medicine & Biology, 1997
- PC-Assisted Dose Assessment in Clinical TL Dosimetry and QA ProgrammesRadiation Protection Dosimetry, 1996
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Dose distributions and dose rate constants for new ytterbium-169 brachytherapy seedsMedical Physics, 1995
- Anisotropy measurements of a high dose rate Ir‐192 source in air and in polystyreneMedical Physics, 1994
- Comparison of radiation dosimetry in water and in solid phantom materials for I‐125 and Pd‐103 brachytherapy sources: EGS4 Monte Carlo studyMedical Physics, 1994
- Volume-based geometric modeling for radiation transport calculationsMedical Physics, 1992
- Dosimetric calculations and measurements of gold plaque ophthalmic irradiators using iridium-192 and iodine-125 seedsInternational Journal of Radiation Oncology*Biology*Physics, 1988
- Specific γ-ray constant and exposure rate constant of 192 IrMedical Physics, 1979