Investigation of the factors responsible for burns during MRI
Open Access
- 21 March 2001
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 13 (4) , 627-631
- https://doi.org/10.1002/jmri.1088
Abstract
Numerous reported burn injuries have been sustained during clinical MRI procedures. The aim of this study was to investigate the possible factors that may be responsible for such burns. Experiments were performed to investigate three possible mechanisms for causing heating in copper wire during MRI: direct electromagnetic induction in a conductive loop, induction in a resonant conducting loop, and electric field resonant coupling with a wire (the antenna effect). Maximum recorded temperature rises were 0.6°C for the loop, 61.1°C for the resonant loop, and 63.5°C for the resonant antenna. These experimental findings suggest that, contrary to common belief, it is unlikely that direct induction in a conductive loop will result in thermal injury. Burn incidents are more likely to occur due to the formation of resonant conducting loops and from extended wires forming resonant antenna. The characteristics of resonance should be considered when formulating safety guidelines. J. Magn. Reson. Imaging 2001;13:627–631.Keywords
This publication has 15 references indexed in Scilit:
- Burns caused by ECG monitoring during MRI imagingAnaesthesia, 1996
- Burns associated with electrocardiographic monitoring during magnetic resonance imagingBurns, 1996
- Burns associated with the use of monitoring equipment during MR proceduresJournal of Magnetic Resonance Imaging, 1996
- Vascular access ports and catheters: Ex vivo testing of ferromagnetism, heating, and artifacts associated with MR imagingMagnetic Resonance Imaging, 1996
- SEVERE BURNS RESULTING FROM MAGNETIC RESONANCE IMAGING WITH CARDIOPULMONARY MONITORINGAmerican Journal of Physical Medicine & Rehabilitation, 1993
- Burn Associated with Temperature Monitoring during Magnetic Resonance ImagingAnesthesiology, 1992
- Burns Associated with Pulse Oximetry during Magnetic Resonance ImagingAnesthesiology, 1991
- Burns associated with clinical MR examinations.Radiology, 1990
- Heating effects of metallic implants by MRI examinationsMagnetic Resonance in Medicine, 1988
- Potential hazards in NMR imaging: heating effects of changing magnetic fields and RF fields on small metallic implantsAmerican Journal of Roentgenology, 1981