Artifacts in camera based single photon emission tomography due to time activity variation
- 1 December 1987
- journal article
- research article
- Published by Springer Nature in European Journal of Nuclear Medicine and Molecular Imaging
- Vol. 13 (9) , 439-442
- https://doi.org/10.1007/bf00281856
Abstract
Image quality in single photon emission computed tomography (SPECT) using a rotating gamma camera is dependent on the time course of the tracer in the field of view. If acquisition times are slow comapred to the tracer turnover, artifacts may occur in the reconstructed images. The properties of such artifacts were studied by computer simulation. Experimental projection data of point sources, cylindrical phantoms, and an anatomically realistic brain phantom were altered by sequentially weighting the projections with a function that varied exponentially or linearly with time. The observed distortion in the reconstructed images could be related to the ratio between the object activity variation and the camera rotation time. If the tracer concentration changed less than a factor of two during one camera rotation then little image distortion was visible although quantitatively the resolution was degraded. If the object's activity variation with time is fast enough to produce noticeable distortion, the artifacts can be reduced by performing multiple, rapid, camera rotations, instead of one rotation, for the same total acquisition time. The proposed procedure is computer storage space intensive and takes longer to produce the transaxial images, but improves image quality.This publication has 4 references indexed in Scilit:
- A study of image discrepancies due to object time-dependence in transmission and emission tomographyPhysics in Medicine & Biology, 1983
- Maximum Likelihood Reconstruction for Emission TomographyIEEE Transactions on Medical Imaging, 1982
- Continuous time-dependence in computed tomographyMedical Physics, 1978
- Principles of computer assisted tomography (CAT) in radiographic and radioisotopic imagingPhysics in Medicine & Biology, 1976