Prospective randomised comparison of irrigated-tip and large-tip catheter ablation of cavotricuspid isthmus-dependent atrial flutter*1

Abstract
Background Radiofrequency (RF) ablation of cavotricuspid isthmus (CTI) dependent flutter can be performed using different types of ablation catheters. It has been proposed that irrigated and large-tip catheters are capable of creating larger lesions, resulting in greater efficacy. This prospective, randomised clinical study compared the efficacy of irrigated and large-tip catheters of different designs. Methods Eighty patients (69 men, 66±11 years) undergoing de novo RF ablation of CTI-dependent flutter were randomised to ablation using one of the following catheters: (i) externally-irrigated \((n=20)\) , (ii) internally-cooled \((n=20)\) , (iii) single sensor, 8-mm tip \((n=20)\) , or (iv) double sensor, 8-mm tip \((n=20)\) . The study endpoint was the demonstration of bidirectional CTI conduction block within 12 min of cumulative RF delivery. Crossover to the externally-irrigated catheter was permitted if this was not achieved. The ablation and procedural parameters, safety and efficacy were compared. Results The primary endpoint was achieved in 64 patients (80%), including all 20 patients randomised to the externally-irrigated catheter. Crossover was required in 16 patients: 9 initially using the internally-cooled catheter (45%), 3 using single-sensor, 8-mm-tip (15%), and 4 using double-sensor, 8-mm-tip (20%) catheters. The higher initial failure rate with the internally-cooled-tip catheter was significant compared to the externally-irrigated \((p=0.001)\) and single-sensor, 8-mm-tip \((p=0.04)\) catheters. The externally-irrigated catheter achieved the study endpoint more frequently with fewer RF applications of shorter duration compared to the internally-cooled-tip catheter and 8-mm-tip catheters, the difference being significant compared with internally cooled ablation. No major complications were observed. Conclusion Among commonly used ablation catheters, the externally-irrigated catheter has a higher efficacy for rapid achievement of CTI block.