A Fast Reconstruction Algorthm for Stationary Positron Emission Tomography Based on a Modified EM Algorithm
- 1 June 1987
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 6 (2) , 98-105
- https://doi.org/10.1109/tmi.1987.4307809
Abstract
An efficient iterative reconstruction method for positron emission tomography (PET) is presented. The algorithm is basically an enhanced EM (expectation maximization) algorithm with improved frequency response. High-frequency components of the ratio of measured to calculated projections are extracted and are taken into account for the iterative correction of image density in such a way that the correction is performed with a uniform efficiency over the image plane and with a flat frequency response. As a result, the convergence speed is not so sensitive to the image pattern or matrix size as the standard EM algorithm, and nonuniformity of the spatial resolution is significantly improved. Nonnegativity of the reconstructed image is preserved. Simulation studies have been made assuming two PET systems: a scanning PET with ideal sampling and a stationary PET with sparse sampling. In the latter, a "bank array" of detectors is employed to improve the sampling in the object plane. The new algorithm provides satisfactory images by two or three iterations starting from a flat image in either case. The behavior of convergence is monitored by evaluating the root mean square of C(b)-1 where C(b) is the correction factor for pixel b in the EM algorithm. The value decreases rapidly and monotonically with iteration number. Although the theory is not accurate enough to assure the stability of convergence, the algorithm is promising to achieve significant saving in computation compared to the standard EM algorithm.Keywords
This publication has 13 references indexed in Scilit:
- Stationary Positron Emission Tomography and Its Image ReconstructionIEEE Transactions on Medical Imaging, 1986
- An Iterative Image Space Reconstruction Algorthm Suitable for Volume ECTIEEE Transactions on Medical Imaging, 1986
- Accelerated Iterative Reconstruction for Positron Emission Tomography Based on the EM Algorithm for Maximum Likelihood EstimationIEEE Transactions on Medical Imaging, 1986
- A Statistical Model for Positron Emission TomographyJournal of the American Statistical Association, 1985
- Maximum Likelihood Reconstruction for Emission TomographyIEEE Transactions on Medical Imaging, 1982
- Sampling Properties of Stationary and Half-Rotation Rings in Positron Emission TomographyJournal of Computer Assisted Tomography, 1981
- A New Sampling Scheme for the Ring Positron Camera: Dichotomic Ring SamplingIEEE Transactions on Nuclear Science, 1981
- Positologica: A Positron ECT Device with a Continuously Rotating Detector RingIEEE Transactions on Nuclear Science, 1980
- "Positology"-the search for suitable detector arrangements for a positron ECT with continuous rotationIEEE Transactions on Nuclear Science, 1979
- A Computer Assisted Ringdector Positron Camera System for Reconstruction Tomography of the BrainIEEE Transactions on Nuclear Science, 1978