On the Life of Self-Quenching Counters
- 15 October 1948
- journal article
- research article
- Published by American Physical Society (APS) in Physical Review B
- Vol. 74 (8) , 898-901
- https://doi.org/10.1103/physrev.74.898
Abstract
A Geiger-Müller counter filled with argon and ethyl acetate was connected to a mass spectrometer, and spectra were obtained after predetermined numbers of counts. Results showed that factors influencing the life of the counter were the disappearance of the quenching vapor and the formation of non-quenching vapors. An argon-methane counter, studied by the same method, showed that contamination of the insides of the counter, by the dissociation products, limited its life. This process occurred before any appreciable fraction of the methane was consumed.Keywords
This publication has 6 references indexed in Scilit:
- Scale Expander for Mass-Spectrometer RecorderReview of Scientific Instruments, 1947
- Adjustable Gas LeakReview of Scientific Instruments, 1947
- On the Role of Polyatomic Gases in Fast CountersPhysical Review B, 1944
- The Factors Influencing the Plateau Characteristics of Self-Quenching Geiger-Mueller CountersPhysical Review B, 1943
- The Discharge Mechanism of Fast G-M Counters from the Deadtime ExperimentPhysical Review B, 1942
- The Discharge Mechanism of Geiger-Mueller CountersPhysical Review B, 1940