Sensitivity and detection limit enhancement of gas chromatographic detection by thermal modulation
- 1 May 1994
- journal article
- research article
- Published by Wiley in Journal of Microcolumn Separations
- Vol. 6 (3) , 229-235
- https://doi.org/10.1002/mcs.1220060306
Abstract
A thermal desorption modulator placed between the outlet of a column and the inlet of a detector enhances chromatographic sensitivity, signal‐to‐noise ratio and detection limit. The modulator operates at a selectable frequency up to 10 Hz, chopping the effluent from the chromatographic column into a series of concentration pulses whose envelope follows the form of the chromatogram. Measuring the difference between the top of each concentration pulse and the following valley demodulates the detector output signal. Thermal modulation increases the recorded signal amplitude because the modulator concentrates substances emerging from the column over a relatively long time into concentration pulses of much shorter duration. Thermal modulation also attenuates low frequency drift. Using this technique enhanced the signal amplitude of a flame ionization detector by a factor of 15 and the signal‐to‐noise ratio by a factor of 5.Keywords
This publication has 7 references indexed in Scilit:
- High-Speed Gas Chromatographic Analysis of a Simulated Process Stream using On-Column Thermal Desorption Modulation for Sample Preconcentration and IntroductionJournal of Chromatographic Science, 1990
- Large-volume sample introduction into narrow-bore gas chromatography columns using thermal desorption modulation and signal averagingJournal of Microcolumn Separations, 1990
- Multiplex gas chromatography by thermal modulation of a fused silica capillary columnAnalytical Chemistry, 1985
- Design and performance of a mass-flow-modulated detector for gas chromatographyJournal of Chromatography A, 1985
- Application of static signal-to-noise theory to the detection and integration of dynamic signalsAnalytica Chimica Acta, 1984
- Solute switching and detection by synchronous demodulation in gas chromatographyJournal of Chromatography A, 1975
- The electron capture detectorJournal of Chromatography A, 1974