Suspected Rotator Cuff Lesions: Tissue Harmonic Imaging versus Conventional US of the Shoulder
- 1 January 2004
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 230 (1) , 243-249
- https://doi.org/10.1148/radiol.2301021517
Abstract
To compare tissue harmonic imaging (THI) of the shoulder with conventional ultrasonography (US) in patients suspected of having rotator cuff lesions. THI and conventional US images were obtained in 50 patients suspected of having rotator cuff lesions. Images were graded for visibility of acromioclavicular (AC) joint surfaces and rotator cuff tendon contours and substances: Grade 1 meant poor; grade 2, moderate; and grade 3, good visibility. Accuracy, sensitivity, and specificity of the diagnosis of AC joint osteoarthritis and accuracy of the diagnosis and rates of underestimation and overestimation of the extent of rotator cuff tears were evaluated. Visibility grade differences were evaluated with the Wilcoxon signed rank test. The McNemar test was used to assess differences in diagnoses. Magnetic resonance (MR) arthrography was the reference standard. Mean visibility grades for readers 1 and 2, respectively, were as follows: for AC joint bone surfaces, 2.1 and 2.0 with THI and 1.7 (P =.010) and 1.7 (P =.16) with conventional US; for AC joint capsule surfaces, 2.2 and 1.9 with THI and 1.8 (P =.005) and 1.8 (P =.34) with US; for supraspinatus tendon contour, 2.6 and 2.2 with THI and 2.1 (P =.001) and 1.9 (P =.055) with US; for supraspinatus tendon substance, 2.2 and 1.9 with THI and 2.0 (P =.036) and 1.7 (P =.070) with US; for subscapularis tendon contour, 2.4 and 2.1 with THI and 2.2 (P =.07) and 2.0 (P =.25) with US; and for subscapularis tendon substance, 1.8 and 1.7 with THI and 2.0 (P =.86) and 1.7 (P =.91) with US. Diagnostic accuracies for the supraspinatus tendon for readers 1 and 2, respectively, were 84% and 74% with THI and 86% and 70% with US (P >.99 for both readers). Corresponding values for the subscapularis tendon were 78% and 72% with THI and 64% (P =.27) and 52% (P =.006) with US. Joint and tendon surface visibility improves with THI, as compared with the visibility achieved with conventional US. THI is superior to conventional US for diagnosis of subscapularis tendon abnormalities.Keywords
This publication has 31 references indexed in Scilit:
- Phase Inversion Tissue Harmonic Sonographic ImagingAmerican Journal of Roentgenology, 2001
- Tissue Harmonic ImagingAmerican Journal of Roentgenology, 2001
- Comparison of Tissue Harmonic Imaging with Conventional US in Abdominal DiseaseRadioGraphics, 2000
- Ultrasonographic evaluation of focal hepatic lesions: comparison of pulse inversion harmonic, tissue harmonic, and conventional imaging techniques.Journal of Ultrasound in Medicine, 2000
- Tissue Harmonic Imaging Sonographie Wird die Bildqualität bei Routine-Ultraschalluntersuchungen im Abdomen verbessert?RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2000
- Clinical use of ultrasound tissue harmonic imagingUltrasound in Medicine & Biology, 1999
- Hepatic sonography: comparison of tissue harmonic and standard sonography techniques.American Journal of Roentgenology, 1999
- Tissue harmonic imaging sonography: evaluation of image quality compared with conventional sonography.American Journal of Roentgenology, 1998
- The development of harmonic distortion in pulsed finite-amplitude ultrasound passing through liverPhysics in Medicine & Biology, 1986
- Prediction of nonlinear acoustic effects at biomedical frequencies and intensitiesUltrasound in Medicine & Biology, 1980