Comparison of Goldmann applanation tonometry with the Tonopen for measuring intraocular pressure in a population-based glaucoma survey in rural West Bengal
- 1 January 2002
- journal article
- research article
- Published by Swets & Zeitlinger Publishers in Ophthalmic Epidemiology
- Vol. 9 (3) , 215-224
- https://doi.org/10.1076/opep.9.3.215.1518
Abstract
Aim This study was undertaken to compare the measurement of intraocular pressure (IOP) in a population-based glaucoma survey by the Goldmann Applanation Tonometer (GAT) and the Tonopen with respect to IOP readings, the time required to measure IOP and subjects' preference for the two instruments. The purpose of the study was to see which of the two instruments could be more suitable in a rural survey, particularly house-to-house, in trying to find cases of advanced glaucoma. method The study was conducted on a total of 203 subjects as a part of a population-based glaucoma survey. The IOP of one randomly selected eye of each subject was measured, first with the GAT and subsequently with the Tonopen. The time taken for each procedure was noted and subjects' preference was ascertained. results The mean of the paired differences between the readings of the two instruments (Tonopen minus GAT) was 1 mm Hg with a standard deviation of 2.28. The Tonopen readings were within ±3 mm Hg of the GAT readings in 86.2% of eyes. In 7.4% of the eyes the paired differences were =±5 mm Hg. Measurement of IOP by the Tonopen was much quicker, the mean time difference being 38.7 sec. Subjects' preference for the Tonopen was higher. conclusion The overall agreement between the measurements of the two instruments was good but a small percentage of large difference (= ±5 mm Hg in 7.4%) may be of concern in a population-based survey. The time to measure IOP and the subjects' preference were strongly in favour of the Tonopen.Keywords
This publication has 11 references indexed in Scilit:
- Examination Methods for Glaucoma Prevalence SurveysArchives of Ophthalmology (1950), 1993
- Evaluation of the Tono-Pen 2 and the X-PertTM Noncontact Tonometers in Cataract SurgeryOphthalmologica, 1993
- Racial Variations in the Prevalence of Primary Open-angle GlaucomaJAMA, 1991
- Microprocessor controlled tonometryEye, 1989
- Accuracy and Precision of the Tono-Pen in Measuring Intraocular Pressure After Keratoplasty and Epikeratophakia and in Scarred CorneasArchives of Ophthalmology (1950), 1988
- The Tono-PenArchives of Ophthalmology (1950), 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Warum ist das selbst mit größter Sorgfalt geeichte Schiötz-Tonometer nicht geeignet den Augendruck zu messen?Klinische Monatsblätter für Augenheilkunde, 1980
- Screening tonometry by techniciansAlbrecht von Graefes Archiv für Ophthalmologie, 1975
- COMPARISON OF SCHIÖTZ AND GOLDMANN TONOMETRY IN A POPULATIONActa Ophthalmologica, 1972