Intensity distribution, modulation transfer function, and the effective dimension of a line‐focus x‐ray focal spot

Abstract
The radiant intensity along the anode-cathode dimension of a line-focus x-ray focal spot tends to peak at the center and decrease gradually to zero at the edges. In recognition of this varying intensity, the ICRU has recommended that the anode-cathode dimension of the focal spot be reported as 0.7 times the dimension measured with a pinhole camera. In this paper more objective correction factors are developed. These correction factors are based upon comparison of calculated MTFs of focal spots with Gaussian and cosine intensity distributions with the MTF for focal spots with a uniform intensity distribution.

This publication has 0 references indexed in Scilit: