A new method of Johnson noise thermometry
- 1 February 1974
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 45 (2) , 151-162
- https://doi.org/10.1063/1.1686578
Abstract
A new method of Johnson noise thermometry was developed and experimentally evaluated. The absolute temperature Ts was determined by measuring the available thermal noise power, i.e., the product of the open‐circuit thermal noise voltage and the short‐circuit thermal noise current generated by a sensing resistor at temperature Ts . The measuredthermal noise power from the thermometer is a linear function of absolute temperature. This new method is independent of the sensing‐resistor composition, the mass and nature of the charge carriers, and, in principle, the ohmic value of the resistor. Consequently, the method has a wide range of application, including use in the realization of an absolute thermodynamic temperature scale and especially in the measurement of high temperatures, particularly in nuclear reactors where the properties of sensormaterials not only change due to the aging effects induced by the high temperatures but also change drastically because of radiation damage and transmutation. An evaluation of the method was performed with the sensor and signal processor both located in an area of high electrical background noise. The signal cable connecting the sensor to the signal processor was about 0.5 m in length. The deviations of the experimental data were not more than ± 0.10% from a straight line calibration through absolute 0 K in the temperature range ΔT from 725 to 1275 K. In the range ΔT, the maximum error in the indicated temperature for a 100% increase in the value of the sensing resistor at a fixed temperature was ± 0.60% and for a 50% decrease, was −1.00%. The feasibility of separating the thermal noise signal from electrical background noise induced in a long signal cable was demonstrated with the sensor and signal processor connected with a 30 m long coaxial cable. With the extraneous noise power in the cable, more than an order‐of‐magnitude greater than the thermal noise power, the uncertainty in temperature due to the extraneous noise was less than 1% at 100 C.Keywords
This publication has 10 references indexed in Scilit:
- Automatic Impedance Measurements Using Thermal Noise AnalysisReview of Scientific Instruments, 1972
- Wide-Band Low-Noise Charge Sensitive PreamplifierIEEE Transactions on Nuclear Science, 1966
- A Low-Noise Charge Sensitive Preamplifier for Semiconductor Detectors Using Paralleled Field-Effect-TransistorsIEEE Transactions on Nuclear Science, 1966
- Thermal Noise in Field-Effect TransistorsProceedings of the IRE, 1962
- A NEW ABSOLUTE NOISE THERMOMETER AT LOW TEMPERATURESCanadian Journal of Physics, 1959
- Low-Temperature Thermal Noise ThermometerReview of Scientific Instruments, 1959
- Measurement of equivalent noise resistance of a noise thermometer amplifierJournal of Scientific Instruments, 1959
- An Absolute Noise Thermometer for High Temperatures and High PressuresReview of Scientific Instruments, 1949
- Thermal Agitation of Electricity in ConductorsPhysical Review B, 1928
- Thermal Agitation of Electric Charge in ConductorsPhysical Review B, 1928