FT-Raman Spectroscopy at 1.339 Micrometers
- 1 June 1994
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 48 (6) , 699-701
- https://doi.org/10.1366/000370294774369009
Abstract
The usual laser employed for Fourier transform Raman spectroscopy is a Nd:YAG unit lasing at 1.064 μm. In this work, use of the 1.339-μm lasing emission from Nd:YAG has been demonstrated. The sensitivity of this instrument is comparable to that of conventional FT-Raman instruments, and excellent anti-Stokes spectra can be easily obtained. Operation further into the near-infrared offers additional possibilities for fluorescence minimization. Results are shown for copper phthalocyanine.Keywords
This publication has 4 references indexed in Scilit:
- Evaluation of a diode laser/charge coupled device spectrometer for near-infrared Raman spectroscopyAnalytical Chemistry, 1989
- Extending the Vibrational Limits in Near-Infrared Fourier Transform Raman SpectroscopyApplied Spectroscopy, 1989
- Fourier transform Raman spectroscopyMicrochimica Acta, 1987
- FT-Raman Spectroscopy: Development and JustificationApplied Spectroscopy, 1986