Comparison of Cardiac MRI on 1.5 and 3.0 Tesla Clinical Whole Body Systems
- 1 July 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 38 (7) , 436-442
- https://doi.org/10.1097/01.rli.0000067489.31556.70
Abstract
A cardiac imaging pilot study was performed on 1.5 and 3.0 Tesla (T) whole body magnetic resonance units equipped with identical gradient sets and geometrically equivalent body coils. The goals were to compare the signal-to-noise (SNR) and contrast-to-noise (CNR) ratios on matched studies conducted at both field strengths and demonstrate the potential for functional and morphologic cardiac evaluation at 3.0 T. Short axis cine true fast imaging with steady precession (True FISP) was compared at 1.5 and 3.0 T using the body coil in transmit-receive mode and transmit-only with single loop and phased array receiver coils. SNR of the myocardium and CNR of the ventricular blood and myocardium were calculated from a quantitative region of interest analysis of these data. Additionally at 3.0 T, long axis and 4-chamber cine as well as “dark blood” imaging are demonstrated with sequence and parameter settings comparable to current state of the art for cardiac evaluation at 1.5 T. The 3.0 T data consistently demonstrates increases in SNR when all imaging conditions are closely matched but the increase has a large variability ranging from 20 to 85% depending on the radiofrequency coil configuration. Ventricular blood–myocardium CNR greater than 30 is obtained at 3.0 T, which is comparable to an optimized 1.5 T acquisition despite the specific absorption rate limitation of flip angle to nearly one half the value. The increased SNR at 3.0 T improves detection of fine anatomic detail, such as the chordae tendineae and mitral valve structure. Increased specific absorption rate can be a limiting fact; however, we have demonstrated that 3.0 T cardiac imaging shows gains in SNR while maintaining the CNR. The SNR gain is advantageous, and phased array coil technology is key for improving cardiac magnetic resonance imaging at 3.0 T.Keywords
This publication has 12 references indexed in Scilit:
- A detunable elliptic transmission line resonator for body imaging at 3 TeslaConcepts in Magnetic Resonance, 2002
- Estimation of heat transfer and temperature rise in partial-body regions during MR procedures: an analytical approach with respect to safety considerationsMagnetic Resonance Imaging, 2002
- Sampling and evaluation of specific absorption rates during patient examinations performed on 1.5-Tesla MR systemsMagnetic Resonance Imaging, 2001
- Cine MR Angiography of the Heart with Segmented True Fast Imaging with Steady-State PrecessionRadiology, 2001
- Magnetization preparation during the steady state: Fat‐saturated 3D TrueFISPMagnetic Resonance in Medicine, 2001
- Human cardiac imaging at 3 T using phased array coilsMagnetic Resonance in Medicine, 2000
- The Intrinsic Signal-to-Noise Ratio in Human Cardiac Imaging at 1.5, 3, and 4 TJournal of Magnetic Resonance, 1997
- "Black blood" T2-weighted inversion-recovery MR imaging of the heart.Radiology, 1996
- The NMR phased arrayMagnetic Resonance in Medicine, 1990
- Radiofrequency penetration and absorption in the human body: Limitations to high‐field whole‐body nuclear magnetic resonance imagingMedical Physics, 1987