Unstable intertrochanteric fractures of the hip. Treatment with Ender pins compared with a compression hip-screw.
- 1 October 1988
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 70 (9) , 1297-1303
- https://doi.org/10.2106/00004623-198870090-00003
Abstract
Operatively were evaluated. The number of secondary operations was much higher in the patients who were treated with Ender pins, and a considerably higher number of those patients had loss of fixation. The use of Ender pins that were too long and a history of a fracture before the operation in which Ender pins were used were statistically significant predictors of secondary operations. In a prospective, consecutive, randomized study of the treatment of unstable intertrochanteric fractures of the hip, the use of Ender pins was compared with the use of the compression hip-screw. There were more than 100 patients in each treatment group. The patients who were alive six months postoperatively were evaluated. The number of secondary operations was much higher in the patients who were treated with Ender pins, and a considerably higher number of those patients had loss of fixation. The use of Ender pins that were too long and a history of a fracture before the operation in which Ender pins were used were statistically significant predictors of secondary operations. Copyright © 1988 by The Journal of Bone and Joint Surgery, Incorporated...This publication has 10 references indexed in Scilit:
- Ender nailing for peritrochanteric fractures of the femur. An analysis of indications, factors related to mechanical failure, and postoperative results.Journal of Bone and Joint Surgery, 1985
- Complications after trochanteric fractures: A comparison between Ender and nail-plate osteosynthesisActa Orthopaedica, 1984
- Complications of Ender-pin fixation in basicervical, intertrochanteric, and subtrochanteric fractures of the hip.Journal of Bone and Joint Surgery, 1983
- Trochanteric and Subtrochanteric Fractures:The Operative Results in a Prospective and Comparative Study of Ender Nailing and McLaughlin OsteosynthesisActa Orthopaedica, 1981
- Pertrochanteric Fractures of the Femur:A Comparative Study of Internal Fixation with Angle Nail-Plates and Flexible Condylar NailsActa Orthopaedica, 1981
- Trochanteric and Subtrochanteric Fractures:One Year Follow-up of a Prospective Study of Ender and McLaughlin OsteosynthesisActa Orthopaedica, 1981
- Ender nailing of intertrochanteric and subtrochanteric fractures of the femur.Journal of Bone and Joint Surgery, 1980
- Unstable Trochanteric Fractures:A Comparative Analysis of Four Methods of Internal FixationActa Orthopaedica, 1980
- Analysis of six hundred and twenty-two intertrochanteric hip fractures.Journal of Bone and Joint Surgery, 1979
- Unstable Trochanteric Fractures Treated with the Sliding Screw-Plate System:A Biomechanical Study of Unstable Trochanteric FracturesIIIActa Orthopaedica, 1978