Design Considerations for Positron Emission Tomography
- 1 February 1981
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-28 (2) , 158-177
- https://doi.org/10.1109/TBME.1981.324788
Abstract
The surge of interest in medical research with positron-emitting nuclides has been accompanied by improvements in the design of positron emission tomographs (scanners). Bismuth germanate has emerged as the scintillation material of choice for high resolution scanners, as its detection efficiency with small crystals is higher than that of NaI. Other detector materials, such as CsF, may be useful when timing accuracy is of great importance, as in high count rate studies or time-of-flight measurements. Circular detector arrays are now generally preferred because they provide the best sensitivity and uniformity of coverage. Multiple rings are used to increase the axial coverage. Other design parameters, such as detector and collimator dimensions, are surveyed and their effect on scanner performance is discussed.Keywords
This publication has 36 references indexed in Scilit:
- POTENTIAL ADVANTAGES OF A CESIUM FLUORIDE SCINTILLATOR FOR A TIME-OF-FLIGHT POSITRON CAMERAJournal of Computer Assisted Tomography, 1980
- Plastic Scintillation Detectors for High Resolution Emission Computed TomographyJournal of Computer Assisted Tomography, 1980
- Positologica: A Positron ECT Device with a Continuously Rotating Detector RingIEEE Transactions on Nuclear Science, 1980
- Semiconductor Gamma-Cameras in Nuclear MedicineIEEE Transactions on Nuclear Science, 1980
- Cesium Fluoride: A New Detector for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1980
- Transverse Section Imaging with the MGH Positron CameraIEEE Transactions on Nuclear Science, 1979
- Design and Performance Characteristics of a Positron Emission Computed Axial Tomograph--ECAT®-IIIEEE Transactions on Nuclear Science, 1979
- Design Considerations for a Positron Emission Transverse Tomograph (PETT V) for Imaging of the BrainJournal of Computer Assisted Tomography, 1978
- Design and Performance Characteristics of a Production Model Positron Imaging SystemIEEE Transactions on Nuclear Science, 1978
- Image Separation Radioisotope ScanningRadiology, 1963