Experimental tests of a single-photon calorimeter for x-ray spectroscopy
- 1 September 1984
- journal article
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 56 (5) , 1263-1266
- https://doi.org/10.1063/1.334130
Abstract
Tests have been made of a nondispersive spectroscopic x-ray detector which operates by measuring the temperature rise following absorption of a single photon. We have observed thermal pulses from 6-keV x-rays, and have resolved the different amplitudes resulting from Mn Kα and K β events. This device was assembled to make quantitative tests of theoretical calculations of the properties of such detectors, and its high heat capacity does not allow it to attain the very high resolution predicted for detectors made by more sophisticated, but still straightforward, techniques. Both the measured resolution of 270-eV full width at half maximum (FWHM) and the absolute amplitude of the response are consistent with predictions. Nonthermal effects in the thermistor limit the precision of this comparison to about 30%.This publication has 4 references indexed in Scilit:
- Thermal detectors as x-ray spectrometersJournal of Applied Physics, 1984
- Bolometer noise: nonequilibrium theoryApplied Optics, 1982
- Measurements of the time-dependent specific heat of amorphous materialsPhysical Review B, 1982
- Non-ohmic conductivity of barely localized electrons in three dimensionsSolid State Communications, 1980