Quantitative PET with positron emitters that emit prompt gamma rays
- 1 January 1995
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 14 (4) , 681-687
- https://doi.org/10.1109/42.476109
Abstract
The purpose of this work was to determine the feasibility of using positron emitting isotopes that emit prompt gammas to acquire quantitative positron emission tomography (PET) data using standard PET instrumentation. Prompt gammas can contaminate PET data by increasing dead time, converting singles into invalid coincidences, and producing multiple coincidences which can lead to the replacement of valid coincidences by invalid coincidences. The measurements in this work were made by scanning point sources containing F-18, Na-22, and Co-60 and studying the effects of the prompt gammas on the PET data, We found that for the Na-22 point source, the annihilation photon coincidence rate was about 25 times the prompt gamma-annihilation photon coincidence rate in the entire active volume of the scanner. With scatter, the Na-22 prompt gamma-annihilation photon coincidence rate was 1.3 times higher than the F-18 scatter coincidence rate. The most significant effect of the prompt gamma was to increase dead time; the dead time correction factor for Cu-60 was 2.4 times higher than the correction factor for N-13 for the same source activity. We conclude that, in many cases, quantitative PET data can be readily obtained with isotopes that emit prompt gammas, using standard PET 2-D instrumentation. However there are some cases, such as 3-D PET, where prompt gammas could significantly contaminate the PET data.Keywords
This publication has 9 references indexed in Scilit:
- Quantitative imaging of I‐124 using positron emission tomography with applications to radioimmunodiagnosis and radioimmunotherapyMedical Physics, 1991
- An evaluation of a two-dimensional array detector for high resolution PETIEEE Transactions on Medical Imaging, 1988
- A Real Time Digital Coincidence Processor for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1986
- EVALUATION AND APPLICATION OF ALUMINA-BASED RB-82 GENERATORS CHARGED WITH HIGH LEVELS OF SR-82/85Journal of Computer Assisted Tomography, 1980
- Trends and Prospects for Circular Ring Positron CamerasIEEE Transactions on Nuclear Science, 1979
- Design Considerations in Positron Computed Tomography (PCT)IEEE Transactions on Nuclear Science, 1979
- LOCALIZATION OF INTRACRANIAL LESIONS BY SCANNING WITH POSITRON-EMITTING ARSENICJAMA, 1955
- The Uses of Nuclear Disintegration in the Diagnosis and Treatment of Brain TumorNew England Journal of Medicine, 1951
- The Use of Positron-emitting Radioisotopes for the Localization of Brain TumorsScience, 1951