A Two-Center Evaluation of the Blood Gas Immediate Response Mobile Analyzer (IRMA)
- 1 January 2002
- journal article
- research article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 40 (2) , 182-191
- https://doi.org/10.1515/cclm.2002.031
Abstract
The Immediate Response Mobile Analyzer (IRMA) is a selective and portable point-of-care testing (POCT) blood gas, electrolyte and hematocrit (Hct) analyzer. The overall analytical performance was evaluated in a two-center study involving two Italian hospital laboratories, following the guidelines suggested by the manufacturer (based on the NCCLS protocol), after a preliminary evaluation of their formal validity. The IRMA was compared to the analyzers used in the routine laboratory as reference. The considered parameters were pH, pO2, pCO2, Na+, K+, ionized calcium and Hct. When using the aqueous quality control material provided by the manufacturer most of the parameters showed good precision, with the exception of pCO2 and pO2 that showed high CVs on two of the three levels of the aqueous control. We could demonstrate that this imprecision was material-related and was reduced when using a different material (blood equilibrated by tonometry). With tonometred blood for pO2 and pCO2 and the aqueous material for the remaining parameters the CVs were all below 5%, ranging from 0.08% to 2.8%. The IRMA results correlated adequately with the comparison instruments, with the exception of sodium and ionized calcium where contradictory results were obtained in the two centers.Keywords
This publication has 8 references indexed in Scilit:
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- Die präklinische BlutgasanalyseDer Anaesthesist, 1998
- Relation between the time to achieve the lower limit of the APTT therapeutic range and recurrent venous thromboembolism during heparin treatment for deep vein thrombosisArchives of internal medicine (1960), 1997
- Clinical experience with the Diametrics IRMA (Immediate Response Mobile Analysis) blood analysis systemPerfusion, 1997
- Accuracy and precision of a new, portable, handheld blood gas analyzer, the IRMA®Journal of Clinical Monitoring and Computing, 1996
- Hand-held blood gas analyzer is accurate in the critical care settingCritical Care Medicine, 1996
- Bedside coagulation monitoring in heparin-treated patients with active thromboembolic disease: A coronary care unit experienceAmerican Heart Journal, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986