Portable tissue equivalent calorimeter

Abstract
A portable tissue equivalent calorimeter has been constructed and employed to measure absorbed dose in a mixed fast-neutron and gamma-ray field. Design details and measurement techniques are described along with the limits of precision and absolute accuracy. Experiments have been carried out with a cyclotron produced neutron beam using the 9Be(d,n)10B reaction at dose rates ranging from 3 to 25 rad/min. A series of ten measurements at a dose rate of approximately 10 rad/min were performed with a precision of +/- 2% (standard deviation about the mean). The compact design of this calorimetric dosimeter system facilitates transport to other laboratories for the measurement of absorbed dose in several complex radiation fields.

This publication has 0 references indexed in Scilit: