Assessment of a microwave‐based clinical waste decontamination unit
- 1 December 1994
- journal article
- Published by Wiley in Journal of Applied Bacteriology
- Vol. 77 (6) , 607-612
- https://doi.org/10.1111/j.1365-2672.1994.tb02808.x
Abstract
P.N. HOFFMAN AND M.J. HANLEY. 1994. A clinical waste decontamination unit that used microwave‐generated heat was assessed for operator safety and efficacy. Tests with loads artificially contaminated with aerosol‐forming particles showed that no particles were detected outside the machine provided the seals and covers were correctly seated. Thermometric measurement of a self‐generated steam decontamination cycle was used to determine the parameters needed to ensure heat disinfection of the waste reception hopper, prior to entry for maintenance or repair. Bacterial and thermometric test pieces were passed through the machine within a full load of clinical waste. These test pieces, designed to represent a worst case situation, were enclosed in aluminium foil to shield them from direct microwave energy. None of the 100 bacterial test pieces yielded growth on culture and all 100 thermal test pieces achieved temperatures in excess of 99d̀C during their passage through the decontamination unit. It was concluded that this method may be used to render safe the bulk of ward‐generated clinical waste.Keywords
This publication has 6 references indexed in Scilit:
- A microbiological study of absorbent padsJournal of Hospital Infection, 1988
- VIRUSES, SPECULA, AND VAGINAL CANCERThe Lancet, 1986
- Susceptibility of hepatitis B virus to disinfectants or heatJournal of Clinical Microbiology, 1984
- Incomplete Inactivation of Hepatitis B Virus after Heat Treatment at 60 C for 10 HoursThe Journal of Infectious Diseases, 1978
- Tests of disinfection by heat in a bedpan washing machineJournal of Clinical Pathology, 1974