Hazardous situation in the MR bore: induction in ECG leads causes fire
- 19 February 2003
- journal article
- case report
- Published by Springer Nature in European Radiology
- Vol. 13 (4) , 690-694
- https://doi.org/10.1007/s00330-003-1841-8
Abstract
There is a potential hazard during examinations of patients with attached or implanted long conductors, e.g. ECG leads: an MR exam of the lumbar spine of a patient was performed in a 1.5-T scanner under ECG monitoring using equipment marked as MR compatible. Standard cabling of 370-cm length was guided without loops from the electrodes through the caudal opening of the magnet bore. During a sagittal T1-weighted turbo-spin-echo scan suddenly a flame of approximately 3 cm arose from the patient's shirt, close to the position of the electrodes. The supervising anaesthesiologist extinguished the flames with his hands. A subsequent physical examination revealed second- to third-degree burns. The analysis of the incident revealed that high voltages can be induced in straight conductors without loops as ECG cables by coupling with the electric component of the HF field. Local heating or sparking can cause an open flame at the position of the electrodes. This danger exists even with ECG equipment that is specifically marked as MR compatible.Keywords
This publication has 10 references indexed in Scilit:
- Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 teslaJournal of Magnetic Resonance Imaging, 2002
- On the heating of linear conductive structures as guide wires and catheters in interventional MRIJournal of Magnetic Resonance Imaging, 2001
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997
- Unusual burns of the lower extremities caused by a closed conducting loop in a patient at MR imaging.Radiology, 1996
- Transesophageal cardiac pacing during magnetic resonance imaging: Feasibility and safety considerationsMagnetic Resonance in Medicine, 1996
- Patient monitoring during clinical MR imaging.Radiology, 1992
- Safety considerations in MR imaging.Radiology, 1990
- Burns associated with clinical MR examinations.Radiology, 1990
- Severe burn of the finger caused by using a pulse oximeter during MR imagingAmerican Journal of Roentgenology, 1989
- The electrical characteristics of some commercial ECG electrodesJournal of Electrocardiology, 1978