Electron beam central axis depth dose measurements
- 1 May 1985
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 12 (3) , 357-361
- https://doi.org/10.1118/1.595696
Abstract
Central axis depth dose measurements were made by the Radiological Physics Center [Texas, USA] on > 70 electron-producing machines used in radiation therapy. These data were consistent for each machine model and nominal energy. Depth dose relations can vary significantly among different machine models for electron beams having the same nominal energy. Both the method used to achieve beam flatness and the mean incident electron energy determine the central axis depth dose curve past the depth of maximum dose. A linear relation of depth dose vs. mean incident electron energy is used to predict depth dose to within 2 mm for most electron beams used clinically at depths greater than d95.This publication has 0 references indexed in Scilit: