Determination of quartz on membrane filters by X-ray diffraction.
Open Access
- 1 December 1979
- journal article
- research article
- Published by Scandinavian Journal of Work, Environment and Health in Scandinavian Journal of Work, Environment & Health
- Vol. 5 (4) , 379-385
- https://doi.org/10.5271/sjweh.2652
Abstract
An X-ray diffraction method is described for the quantitative determination of quartz [an air pollutant] in dust samples. The analysis was done directly on a collecting cellulose ester membrane filter of the type used in an air sampler. Quantitation was achieved by the measurement of the quartz line intensities with a digital integrator and by the comparison of these values with those measured from calibration standards prepared on similar filters with water suspensions. The overall precision of the procedure was 8%. When the method was applied to foundry dusts, the accuracy of the quartz analysis was found to be within 10% at the 1-10-mg level of total loading. The detection limit for quartz was below 0.1 mg.This publication has 0 references indexed in Scilit: