Low-Level Liquid Scintillation Counting in An Underground Laboratory
- 1 January 1980
- journal article
- research article
- Published by Cambridge University Press (CUP) in Radiocarbon
- Vol. 22 (2) , 505-511
- https://doi.org/10.1017/s0033822200009826
Abstract
A commercial liquid scintillation counter (LSC) was tested for its ability of low level counting (LLC) in two underground laboratories (70m of water equivalent cover). Reduction of total background and background components compared to the normal counting laboratory (5m w.e. cover) up to a factor of 9 depending on the energy range and the shielding of counting cell and laboratory walls were obtained. First experiments with a large volume cell (50cc) led to a modern 14C net effect of 425cpm and a corresponding background of 1.6cpm.Keywords
This publication has 2 references indexed in Scilit:
- Low-Level Gas Proportional Counting in An Underground LaboratoryRadiocarbon, 1980