Estimation of aortic compliance using magnetic resonance pulse wave velocity measurement
- 19 May 2000
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (6) , 1703-1713
- https://doi.org/10.1088/0031-9155/45/6/320
Abstract
A method for compliance estimation employing magnetic resonance pulse wave velocity measurement is presented. Time-resolved flow waves are recorded at several positions along the vessel using a phase contrast sequence, and pulse wave velocity is calculated from the delay of the wave onsets. Using retrospective cardiac gating in combination with an optically decoupled electrocardiogram acquisition, a high temporal resolution of 3 ms can be achieved. A phantom set-up for the simulation of pulsatile flow in a compliant vessel is described. In the phantom, relative errors of pulse wave velocity estimation were found to be about 15%, whereas in a volunteer, larger errors were found that might be caused by vessel branches. Results of pulse wave velocity estimation agree with direct aortic distension measurements which rely on a peripheral estimate of aortic pressure and are therefore less accurate. Studies in 12 volunteers show values of pulse wave velocity consistent with the literature; in particular the well-known increase in pulse wave velocity with age was observed. Preliminary results show that the method can be applied to aortic aneurysms.Keywords
This publication has 16 references indexed in Scilit:
- Accuracy of arterial pulse‐wave velocity measurement using MRJournal of Magnetic Resonance Imaging, 1998
- Direct Magnetic Resonance Determination of Aortic Distensibility in Essential HypertensionHypertension, 1997
- Measurement of regional aortic compliance by MR imaging: A study of reproducibilityJournal of Magnetic Resonance Imaging, 1995
- A velocity correlation method for measuring vascular compliance using MR imagingJournal of Magnetic Resonance Imaging, 1995
- Synchronization device for electrocardiography-gated echo-planar imaging.Radiology, 1995
- Pulsewave velocity measurement using a new real-time MR-methodMagnetic Resonance Imaging, 1995
- Ultrafast interleaved gradient‐echo‐planar imaging on a standard scannerMagnetic Resonance in Medicine, 1993
- Comparison between retrospective gating and ECG triggering in magnetic resonance velocity mappingMagnetic Resonance Imaging, 1993
- Aortic compliance measured by non-invasive Doppler ultrasound: Description of a method and its reproducibilityClinical Science, 1990
- Blood Flow Imaging by Cine Magnetic ResonanceJournal of Computer Assisted Tomography, 1986