Production of arrays of chemically distinct nanolitre plugs via repeated splitting in microfluidic devices
- 27 July 2006
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 6 (9) , 1178-1186
- https://doi.org/10.1039/b604993a
Abstract
This paper reports a method for the production of arrays of nanolitre plugs with distinct chemical compositions. One of the primary constraints on the use of plug-based microfluidics for large scale biological screening is the difficulty of fabricating arrays of chemically distinct plugs on the nanolitre scale. Here, using microfluidic devices with several T-junctions linked in series, a single input array of large (∼320 nL) plugs was split to produce 16 output arrays of smaller (∼20 nL) plugs; the composition and configuration of these arrays were identical to that of the input. This paper shows how the passive break-up of plugs in T-junction microchannel geometries can be used to produce a set of smaller-volume output arrays useful for chemical screening from a single large-volume array. A simple theoretical description is presented to describe splitting as a function of the Capillary number, the capillary pressure, the total pressure difference across the channel, and the geometric fluidic resistance. By accounting for these considerations, plug coalescence and plug–plug contamination can be eliminated from the splitting process and the symmetry of splitting can be preserved. Furthermore, single-outlet splitting devices were implemented with both valve- and volume-based methods for coordinating the release of output arrays. Arrays of plugs containing commercial sparse matrix screens were obtained from the presented splitting method and these arrays were used in protein crystallization trials. The techniques presented in this paper may facilitate the implementation of high-throughput chemical and biological screening.Keywords
This publication has 23 references indexed in Scilit:
- Microgram-Scale Testing of Reaction Conditions in Solution Using Nanoliter Plugs in Microfluidics with Detection by MALDI-MSJournal of the American Chemical Society, 2006
- Formation of droplets and bubbles in a microfluidic T-junction—scaling and mechanism of break-upLab on a Chip, 2006
- Torque-Actuated Valves for MicrofluidicsAnalytical Chemistry, 2005
- Controlling Nonspecific Protein Adsorption in a Plug-Based Microfluidic System by Controlling Interfacial Chemistry Using Fluorous-Phase SurfactantsAnalytical Chemistry, 2004
- Reagent-Loaded Cartridges for Valveless and Automated Fluid Delivery in Microfluidic DevicesAnalytical Chemistry, 2004
- Screening of Protein Crystallization Conditions on a Microfluidic Chip Using Nanoliter-Size DropletsJournal of the American Chemical Society, 2003
- Rapid Prototyping of Microfluidic Systems in Poly(dimethylsiloxane)Analytical Chemistry, 1998
- Dynamics of a drop in a constricted capillary tubeJournal of Fluid Mechanics, 1994
- Effect of surfactants on the motion of drops through circular tubesPhysics of Fluids A: Fluid Dynamics, 1992