Pressure Injection on a Valved Microdevice for Electrophoretic Analysis of Submicroliter Samples
- 16 April 2005
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 77 (11) , 3637-3643
- https://doi.org/10.1021/ac048115z
Abstract
A recent report describes a reversible valve that can be used in series to achieve diaphragm pumping on chip (Grover, W. H.; Skelley, A. M.; Liu, C. N.; Lagally, E. T.; and Mathies, R. A. Sens. Actuators, B 2003, 89, 315−323). Here, the functionality of an integrated diaphragm pump on a hybrid PDMS−glass microchip to perform pressure injections for electrophoretic separations is demonstrated. A chip design that can perform both pressure and electrokinetic (EK) injection is described, and a mixture of fluorescein and ROX dyes in borate buffer is utilized as a model sample system. Multiple electrophoretic separations of sample injected with pressure and voltage are compared. Over multiple EK injections, an electrophoretic bias is observed and the injected analytes are not representative of the sample, with the peak area ratio changing 20% after 20 runs. Over multiple pressure injections, however, the sample composition is maintained, with a 3.6% CV over 20 runs. The data presented show the ability to alternate between injection types and pressure-inject a representative sample volume after a bias has already been observed with multiple EK injections. Multiple pressure injections have been performed on sample volumes as low as 500 nL while maintaining sample composition, supporting its use in integrated systems for small-volume sampling.Keywords
This publication has 21 references indexed in Scilit:
- Extending the Lifetime of the Running Electrolyte in Capillary Electrophoresis by Using Additional Compartments for External ElectrolysisAnalytical Chemistry, 2004
- Micro Total Analysis Systems. Recent DevelopmentsAnalytical Chemistry, 2004
- Sampling BIAS at Channel Junctions in Gated Flow Injection on ChipsAnalytical Chemistry, 2002
- Micro Total Analysis Systems. 1. Introduction, Theory, and TechnologyAnalytical Chemistry, 2002
- Monolithic Microfabricated Valves and Pumps by Multilayer Soft LithographyScience, 2000
- Dynamic On-Column pH Monitoring in Capillary Electrophoresis: Application to Volume-Limited Outlet VialsAnalytical Chemistry, 1996
- Artifacts Related to Sample Introduction in Capillary Gel Electrophoresis Affecting Separation Performance and QuantitationAnalytical Chemistry, 1995
- Glass chips for high-speed capillary electrophoresis separations with submicrometer plate heightsAnalytical Chemistry, 1993
- Capillary electrophoresis and sample injection systems integrated on a planar glass chipAnalytical Chemistry, 1992
- Effects of the sample matrix on the separation of peptides by high performance capillary electrophoresisJournal of High Resolution Chromatography, 1991