Electrical impedance tomography with compensation for electrode positioning variations
- 1 June 1998
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 43 (6) , 1731-1739
- https://doi.org/10.1088/0031-9155/43/6/025
Abstract
Ideally electrical impedance tomography (EIT) should not be oversensitive to electrode positions, but this conflicts with efforts to produce high-resolution images. Two procedures are presented that balance reducing the sensitivity to electrode position errors with generating practicable EIT images. The first provides a criterion based on electrode sensitivity for regularizing the reconstruction through spectral expansion. The main consequences of this are that smoother images are produced and the number of artefacts and their magnitude are generally reduced. The second modification uses the recorded data to compensate for electrode movements that have occurred after the reference data were measured. Image smoothness is used as the criterion for the readjustment. Computer simulation tests have shown that this modification produces improved image fidelity.Keywords
This publication has 5 references indexed in Scilit:
- Impedance imaging of lung ventilation: do we need to account for chest expansion?IEEE Transactions on Biomedical Engineering, 1996
- Determination of the boundary of an object inserted into a water-filled cylinderClinical Physics and Physiological Measurement, 1992
- Spectral expansion analysis in electrical impedance tomographyJournal of Physics D: Applied Physics, 1991
- Errors in reconstruction of resistivity images using a linear reconstruction techniqueClinical Physics and Physiological Measurement, 1988
- Singular value decomposition and least squares solutionsNumerische Mathematik, 1970