The Mark IV System for Radionuclide Computed Tomography of the Brain
- 1 November 1976
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 121 (2) , 405-413
- https://doi.org/10.1148/121.2.405
Abstract
The Mark IV scanning system is a simple four-sided arrangement of 32 independent detectors which rotate continuously as a unit, detecting, processing, and displaying the reconstructed data while the study progresses. Detection is by single photon counting and is compatible with commercially available radionuclides. An empirical correction is applied for attenuation, difference in detector response, and scatter. It is a high-sensitivity device with approximately uniform resolution throughout the section plane. There is good reproducibility and accuracy for absolute quantification of radionuclide concentration in the brain. Clinical applications include scans of 99mTcO4, 99mTc-RBC, 123I-iodoantipyrine, 99mTc-diphosphonate, and 111In-DTPA.Keywords
This publication has 3 references indexed in Scilit:
- DESIGN AND PERFORMANCE-CHARACTERISTICS OF A WHOLE-BODY POSITRON TRANSAXIAL TOMOGRAPH1976
- Cylindrical and Section Radioisotope Scanning of the Liver and BrainRadiology, 1964
- Regional Blood Flow by Fractional Distribution of IndicatorsAmerican Journal of Physiology-Legacy Content, 1958