A new “transducer‐tipped” fiber optic catheter for measuring intramuscular pressures
- 1 May 1990
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 8 (3) , 464-468
- https://doi.org/10.1002/jor.1100080318
Abstract
Laboratory and clinical tests were used to determine the efficacy of a new fiber optic “transducer-tipped” catheter for measuring intramuscular pressures. When pressures ranging from 0 to 250 mm Hg were applied by a mercury manometer, the fiber optic system accurtely recorded the pressures. In addition, the fiber optic system showed long-term stability by accurately recording the manometer pressure over a 3 day period. The Slit system showed a higher magnitude of hydrostatic pressure artifacts with catheter tip movement as compared to the fiber optic system. The two catheters showed no difference when measuring pressures in pig muscle at rest or when being compressed throughout a range of 0 to 250 mm Hg. In human volunteers, both catheters measured essentially equal pressures at rest, during venous stasis, and during a combination of venous stasis and compression. For long-term assessment, the Slit system required as many as three saline flushes, whereas the fiber optic system measured pressures continuously without manipulation. We conclude that the fiber optic system is as accurate as the Slit catheter for measuring tissue fluid pressures at rest. In addition, the fiber optic system offers distinct advantages over conventional fluid-filled systems for measuring intramuscular pressures due to a lack of hydrostatic pressure artifacts caused by limb position and to the lack of flushing for long-term measurements.Keywords
This publication has 8 references indexed in Scilit:
- Intraarticular pressure during continuous passive motion of the human kneeJournal of Orthopaedic Research, 1989
- Continuous monitoring of intracranial pressure with a miniaturized fiberoptic deviceJournal of Neurosurgery, 1987
- Myocardial protection in the neonatal heartThe Journal of Thoracic and Cardiovascular Surgery, 1987
- Diagnosis of chronic anterior compartment syndrome in the lower legActa Orthopaedica, 1987
- Monitoring dynamic anterior compartment pressures during exerciseThe American Journal of Sports Medicine, 1982
- COMPARTMENTAL PRESSURE MEASUREMENTS - AN EXPERIMENTAL INVESTIGATION USING THE SLIT CATHETER1981
- Reevaluation of the needle method for measuring interstitial fluid pressureAmerican Journal of Physiology-Legacy Content, 1975
- Pressure‐Volume Characteristics of the Interstitial Fluid Space in the Skeletal Muscle of the CatActa Physiologica Scandinavica, 1974