Background. Mechanically ventilated very low birth weight infants often present with frequent episodes of hypoxemia, and maintaining arterial oxygen saturation by pulse oximetry (Spo2) within a normal range by manual fraction of inspired oxygen (Fio2) adjustments is difficult and time consuming. Objectives. An algorithm for closed-loop Fio2 control (cFio2) to maintain Spo2 within a target range was compared with continuous manual Fio2(mFio2) adjustments by a nurse in a group of ventilated infants who presented with frequent episodes of hypoxemia. Results. Fourteen infants (birth weight: 712 ± 142 g; gestational age: 25 ± 1.6 weeks; age: 26 ± 11 days; synchronized intermittent mandatory ventilation rate: 24 ± 10 b/m; peak inspiratory pressure: 17.5 ± 2.0 cmH2O; positive end-expiratory pressure: 4.3 ± 0.5 cmH2O) were studied for 2 hours on each mode in random sequence. Both modes aimed to maintain Spo2 between 88% and 96%. There were 15 ± 7 and 16 ± 6 hypoxemic episodes/hour (Spo2 5 s) during mFio2 and cFio2, respectively; episode duration was 41 ± 23 and 32 ± 15 s, totaling 19 ± 16% and 17 ± 12% of recording time. There were 13 ± 10 and 10 ± 8 hyperoxemic episodes/hour (Spo2>96%, >5 s) during mFio2 and cFio2,respectively; episode duration was 27 ± 15 and 24 ± 19 s, totaling 15 ± 14% and 10 ± 9% of recording time. Mean Spo2 and Fio2 levels were similar during both modes. The nurse made 29 ± 17 adjustments/hour during mFio2. There was a significant increase in the duration of normoxemia (Spo2 between 88%–96%) during cFio2 (75 ± 13 vs 66 ± 14% of recording time). Conclusion. In this group of infants, cFio2 was at least as effective as a fully dedicated nurse in maintaining Spo2 within the target range, and it may be more effective than a nurse working under routine conditions. We speculate that during long-term use, cFio2 may save nursing time and reduce the risks of morbidity associated with supplemental oxygen and episodes of hypo- and hyperoxemia.Keywords
This publication has 25 references indexed in Scilit:
- Prone positioning decreases episodes of hypoxemia in extremely low birth weight infants (1000 grams or less) with chronic lung diseaseThe Journal of Pediatrics, 1997
- Mechanisms for episodes of hypoxemia in preterm infants undergoing mechanical ventilationThe Journal of Pediatrics, 1995
- Episodes of spontaneous desaturations in infants with chronic lung disease at two different levels of oxygenationPediatric Pulmonology, 1993
- Adaptive control of inspired oxygen delivery to the neonatePediatric Pulmonology, 1992
- Spontaneous desaturations in intubated very low birth weight infants with acute and chronic lung diseasePediatric Pulmonology, 1992
- A Cohort Study of Transcutaneous Oxygen Tension and the Incidence and Severity of Retinopathy of PrematurityNew England Journal of Medicine, 1992
- Differential effects of oxygen and barotrauma on lung injury in the neonatal pigletPediatric Pulmonology, 1991
- Oxygen Toxicity in the Neonatal PeriodActa Paediatrica, 1990
- Hypoxic airway constriction in infants of very low birth weight recovering from moderate to severe bronchopulmonary dysplasiaThe Journal of Pediatrics, 1989
- New technique for servo-control of arterial oxygen tension in preterm infants.Archives of Disease in Childhood, 1979