Measurement of dose distributions using film in therapeutic electron beams
- 1 November 1989
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 16 (6) , 911-915
- https://doi.org/10.1118/1.596409
Abstract
The feasibility of using film dosimetry data as the input data for patient treatment planning was evaluated. The central-axis depth dose and the off-axis ratios obtained from film measurements in a solid phantom were compared with those of ion-chamber measurements in water. Two techniques were used to generate isodose distributions. The first technique used only the film data, i.e., the central-axis depth dose and the off-axis ratios used for the reconstruction were determined from the film optical density (corrected for film nonlinearity). In the second technique, the central-axis depth dose measured by an ion chamber in a water phantom was combined with the off-axis ratios measured using film in the ''''solid water'''' phantom. The resulting isodose distributions from both techniques were compared with the ion-chamber measurements in water for 7-, 12-, and 18-MeV electrons, and the second technique showed better agreement with the ion-chamber measurements than did the first technique. The differences were within a clinically acceptable range.This publication has 3 references indexed in Scilit: