A noninvasive thermoplastic head immobilization system
- 1 June 1995
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 82 (6) , 1082-1085
- https://doi.org/10.3171/jns.1995.82.6.1082
Abstract
✓ The authors have developed a noninvasive head immobilization system for use in neuroimaging (magnetic resonance imaging, computerized tomography, single photon emission computerized tomography, and projection angiography), neurosurgical planning, and neurosurgery. These diagnostic and surgical procedures require patient immobilization, reproducible patient positioning, and anatomical localization. The thermoplastic system described in this technical note addresses each of these requirements with a high degree of accuracy and with no bone fixation. The reproducibility of positioning and effectiveness of immobilization were evaluated using nine healthy volunteers during repeated sessions of magnetic resonance imaging. The mean axial displacement for repeated positioning was 0.6 mm (variance 0.1 mm); the mean displacement during robust patient motion in the axial direction was 1.8 mm (variance 0.9 mm).Keywords
This publication has 7 references indexed in Scilit:
- The Bremen vacuum cap — a valuable positioning device for use in MRIMagnetic Resonance Materials in Physics, Biology and Medicine, 1993
- An Independent Application Accuracy Evaluation of Stereotactic Frame SystemsStereotactic and Functional Neurosurgery, 1992
- A new model for non-invasive, reproducible fixation of a stereotaxic frame using an orthodontic resin plateActa Neurochirurgica, 1992
- An afterloading brachytherapy device utilizing thermoplastic materialRadiotherapy and Oncology, 1989
- Noninvasive multipurpose stereoadapterNeurological Research, 1987