Determination of Saccharin in Plasma and Urine by High Performance Liquid Chromatography
- 1 January 1980
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 13 (10) , 805-816
- https://doi.org/10.1080/00032718008078002
Abstract
A sensitive and specific high performance liquid chromatographic (HPLC) assay for the determination of saccharin in plasma and urine was developed. Saccharin is extracted into diethyl ether at acid pH, evaporated, and reconstituted prior to instrumental analysis. Overall recovery of saccharin is 86.9 + 8.6% and the sensitivity limits of detection is 0.15 μg per ml of plasma or urine using a fluorescence detector. The sensitivity limit in plasma can be extended to 20 ng per ml by use of a 2 ml assay volume and detector attenuation. The assay was used for the determination of saccharin in plasma and urine of rats following oral doses of 5 mg/kg.Keywords
This publication has 4 references indexed in Scilit:
- High‐Pressure Liquid Chromatographic Determination of Saccharin in Artificial Sweeteners and PharmaceuticalsJournal of Pharmaceutical Sciences, 1979
- Absorption of Saccharin from Rat Urinary BladderJournal of Pharmaceutical Sciences, 1978
- Determination of saccharin in urine by electron-capture gas chromatography after extractiv methylationJournal of Chromatography A, 1978
- Saccharin. Distribution and excretion of a limited dose in the ratJournal of Agricultural and Food Chemistry, 1973