Endotracheal Tube Safety During Electrodissection Tonsillectomy
- 1 June 1992
- journal article
- case report
- Published by American Medical Association (AMA) in JAMA Otolaryngology–Head & Neck Surgery
- Vol. 118 (6) , 643-645
- https://doi.org/10.1001/archotol.1992.01880060093019
Abstract
• A case report of an endotracheal tube fire occurring during electrodissection tonsillectomy is presented. The authors believe that this incident occurred because a retrograde leak of ventilating gases around an uncuffed endotracheal tube during positive-pressure ventilation produced a high oxygen concentration in the mouth, allowing indirect ignition of the tube. In vitro testing supported this hypothesis. Ignition tests on polyvinylchloride endotracheal tubes using electrocautery in various oxygen concentrations were performed. As oxygen concentration increased, the endotracheal tube could be moved further from the cautery and still allow ignition of the tube. At 52% oxygen, with the cautery set at 25-W coagulation current, the endotracheal tube could not be ignited. Recommendations to prevent a recurrence of this incident are included. (Arch Otolaryngol Head Neck Surg.1992;118:643-645)Keywords
This publication has 7 references indexed in Scilit:
- Electrocautery-Induced Airway Fire During TracheostomyAnesthesia & Analgesia, 1990
- Electrodissection TonsillectomyJAMA Otolaryngology–Head & Neck Surgery, 1990
- Flammability of Endotracheal Tubes in Oxygen and Nitrous Oxide Enriched AtmosphereAnesthesiology, 1987
- Endotracheal Tube Fire Ignited by Pharyngeal ElectrocauteryAnesthesiology, 1986
- Helium Retards Endotracheal Tube Fires from Carbon Dioxide LasersAnesthesiology, 1985
- Endotracheal Tube Ignition during Laryngeal Surgery with ResectoscopeAnesthesiology, 1982
- Indirect Ignition of the Endotracheal Tube During Carbon Dioxide Laser SurgeryJAMA Otolaryngology–Head & Neck Surgery, 1980