Prototype of a Fully Automated Device for Determination of Bacterial Antibiotic Susceptibility in the Clinical Laboratory 1
Open Access
- 1 January 1971
- journal article
- research article
- Published by American Society for Microbiology in Applied Microbiology
- Vol. 22 (6) , 980-986
- https://doi.org/10.1128/aem.22.6.980-986.1971
Abstract
A completely automated system for the performance of antibiotic susceptibility tests in the clinical laboratory is described. With a modicum of personnel involvement, data on 40 specimens tested against 13 antibiotics are obtained every hour after an initial 3-hr period. The step by step explanation of the functioning of this prototype system, based on a thoroughly tested manual model, is presented. The system compares well with the standard diffusion test and has a potential for application to other endeavors of the clinical microbiology laboratory with a comparable saving in time and labor. ImagesKeywords