Ventilation/perfusion indices do not correlate with the difference between oxygen consumption measured by the Fick principle and metabolic monitoring systems in critically ill patients
- 1 April 1992
- journal article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 20 (4) , 479-482
- https://doi.org/10.1097/00003246-199204000-00008
Abstract
To determine whether the difference between oxygen consumption (VO2) measured by metabolic gas monitoring systems and by the Fick principle is related to venous admixture, deadspace/tidal volume ratio, or alveolar-arterial oxygen tension gradient in critically ill patients. A prospective study. An 11-bed general ICU in a 900-bed teaching hospital. Twenty critically ill patients admitted to the ICU who required mechanical ventilation, right heart catheterization, and arterial and mixed venous gas measurements for normal clinical management. Thirty-three recordings were analyzed. The mean VO2 measured by the metabolic gas monitoring system was 308 +/- 63.9 (SD) mL/min and was significantly greater than the mean VO2 measured by the Fick principle of 284 +/- 72.0 mL/min. The difference between the two measurements of 24.3 +/- 47.6 mL/min correlated poorly with venous admixture (r2 = .0009), dead-space/tidal volume ratio (r2 = .0064) and alveolar-arterial oxygen tension gradient (r2 = .017). If the difference in VO2 measured by metabolic gas monitoring systems and the Fick principle is due to intrapulmonary VO2 then in critically ill patients the ventilation/perfusion indices of venous admixture, deadspace/tidal volume ratio and alveolar-arterial oxygen tension gradient correlate poorly with intrapulmonary VO2.Keywords
This publication has 0 references indexed in Scilit: