A Cyclone for Size-Selective Sampling of Ambient Air
Open Access
- 1 August 1980
- journal article
- research article
- Published by Taylor & Francis in Journal of the Air Pollution Control Association
- Vol. 30 (8) , 872-876
- https://doi.org/10.1080/00022470.1980.10465122
Abstract
A cyclone with a 47 mm after-filter has been developed for ambient air size-selective monitoring. It has been extensively evaluated with laboratory-generated aerosol. Variation of the pressure drop and 50% cut point with flow rate show that the cyclone operates in a single flow regime with a vortex in the outlet flow. The particle size cutoff curve is comparable in sharpness to a cascade impactor and is the same for solid or liquid particles. At 21.7 L/min, D 50 is 2.5μm and at 15.4 L/min, D 50 is 3.5 μm. Collection efficiency data for flow rates from 8 to 27 L/min fit a universal curve when plotted vs. the normalized particle diameter, (D-D 50)/D 50 Reentrainment of previously deposited particles is less than 1 % of the loading per day. In field tests the cyclone has proved to be a very satisfactory size-selective sampler.This publication has 1 reference indexed in Scilit: