Comparison of rebiopsy rates after stereotactic core needle biopsy of the breast with 11-gauge vacuum suction probe versus 14-gauge needle and automatic gun.
- 1 March 1999
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 172 (3) , 683-687
- https://doi.org/10.2214/ajr.172.3.10063860
Abstract
The 11-gauge vacuum suction probe is an alternative to the 14-gauge needle and automatic gun for performing stereotactic core needle biopsies. This study compares rebiopsy rates after stereotactic core needle biopsies that were performed with the two methods. The study also assesses the outcomes of those repeat biopsies.Five hundred ninety-two stereotactic core needle biopsies using a 14-gauge needle and automatic gun and 354 using an 11-gauge vacuum suction probe were performed consecutively. Excluding malignancies, the number of cases requiring rebiopsy and the reasons for rebiopsy were determined for each group. The histologic diagnoses of the repeat biopsies were assessed.The rebiopsy rate was significantly lower with the 11-gauge vacuum suction probe (9.0%) than with the 14-gauge needle and automatic gun (14.9%) (p = .013). Significant reductions were found in cases of insufficient sampling (probe, 1.7%; needle, 4.4%; p = .042) and mammographic-pathologic discrepancy (probe, 0.8%; needle, 3.4%; p = ....Keywords