Evaluation of the Keeler Pulsair non‐contact tonometer
- 1 October 1989
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 67 (5) , 567-572
- https://doi.org/10.1111/j.1755-3768.1989.tb04109.x
Abstract
The recently-introduced Pulsair non-contact tonometer (Keeler Instruments, Inc. Broomall, PA [USA]) was evaluated against the Goldmann applanation tonometer in cannulated post mortem human eyes, and in living subjects at three clinical centers. Statistical analysis of the findings revealed strong linear relationships between Goldmann or manometric and Pulsair intraocular pressure readings, with correlation coefficients ranging from 0.79 to 0.97. However, with the Pulsair reading as the dependent variable, the slope of the relationship in each case was significantly less than 1.0 (range 0.65 to 0.83), while the intercept was positive (range 1.31 to 2.33 mmHg). Thus, the Pulsair tended to read low at intraocular pressure above the normal range. Conversion factors and a calibration table allowing intraocular pressure estimates more comparable to Goldmann tonometry using the current (J-series) Pulsair are provided herein. A calibration algorithm could be integrated into the circuitry of future production models of the Pulsair instrument.Keywords
This publication has 6 references indexed in Scilit:
- Comparison of a new non-contact tonometer with Goldmann applanationEye, 1989
- A new handheld air impulse tonometerEye, 1988
- A Noncontact Applanation TonometerArchives of Ophthalmology (1950), 1974
- SIMULTANEOUS MEASUREMENT OF CHANGING INTRAOCULAR PRESSURE AND OUTFLOW FACILITY IN THE VERVET MONKEY BY CONSTANT PRESSURE INFUSION1964
- The Goldmann Applanation Tonometer*American Journal of Ophthalmology, 1958
- Zur Applanationstonometrie an der SpaltlampeOphthalmologica, 1957