Apparent Loss of Urinary Albumin during Long-term Frozen Storage: HPLC vs Immunonephelometry
Open Access
- 1 August 2007
- journal article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 53 (8) , 1520-1526
- https://doi.org/10.1373/clinchem.2007.088823
Abstract
Background: Urinary albumin detection by immunonephelometry is decreased by ∼30% in samples that have been frozen at −20 °C. An HPLC method for assessment of urinary albumin that detects immunoreactive and immunochemically nonreactive albumin has been introduced as an alternative to immunonephelometry. We investigated whether this technique is affected by sample temperature, particularly freezing. Methods: Urine samples (n = 295) were collected from the general population (Prevention of Renal and Vascular End-Stage Disease Study). Samples were assessed by both immunonephelometry and HPLC when fresh and after storage at −20 °C for 4, 8, and 12 months and at −80 °C for 12 months. Results: With immunonephelometry, storage for 4, 8, and 12 months at −20 °C resulted in mean (SD) urine albumin changes of −21% (29%), −28% (29%), and −34% (31) (P <0.001 for trend). Storage at −80 °C resulted in a 5% (19%) change after 12 months of storage (not significant). With HPLC, storage for 4, 8, and 12 months at −20 °C resulted in urine albumin changes of −33% (28%), −43% (24%), and −55% (21%; P <0.001 vs immunonephelometry). Storage at −80 °C resulted in a −29% (29%) change (P <0.001 vs immunonephelometry). Conclusion: Loss of albumin after freezing urine depends not only on freezing temperature but also on detection method. Detection of albumin by immunonephelometry appears to be significantly less influenced by freezing than detection by HPLC. Storage at −80 °C appears to prevent loss when using immunonephelometry, whereas HPLC still shows considerable loss even when urine is frozen at −80 °C. We propose that for reliable measurement of urine albumin, fresh samples should be used.Keywords
Funding Information
- Dutch Kidney Foundation (E.013)
This publication has 16 references indexed in Scilit:
- Prolonged Frozen Storage of Urine Reduces the Value of Albuminuria for Mortality PredictionClinical Chemistry, 2007
- Analytical validation of an HPLC assay for urinary albuminClinica Chimica Acta; International Journal of Clinical Chemistry, 2006
- The analysis and characterisation of immuno-unreactive urinary albumin in healthy volunteersClinical Biochemistry, 2006
- Urinary Albumin Excretion as a Predictor of the Development of Hypertension in the General PopulationJournal of the American Society of Nephrology, 2006
- Falsely Low Urinary Albumin Concentrations after Prolonged Frozen Storage of Urine SamplesClinical Chemistry, 2005
- Performance Characteristics of an HPLC Assay for Urinary AlbuminAmerican Journal of Clinical Pathology, 2005
- Characterization of Immunochemically Nonreactive Urinary AlbuminClinical Chemistry, 2004
- Which method for quantifying urinary albumin excretion gives what outcome? A comparison of immunonephelometry with HPLCKidney International, 2004
- Earlier detection of microalbuminuria in diabetic patients using a new urinary albumin assayKidney International, 2004
- Freezing method affects the concentration and variability of urine proteins and the interpretation of data on microalbuminuriaDiabetic Medicine, 2000