Cobalt‐60 dose calibration: A comparison of methods
- 1 November 1979
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 6 (6) , 530-535
- https://doi.org/10.1118/1.594645
Abstract
Cobalt-60 dose determination can be accomplished by a variety of methods. This study investigates dose determination by actual "at-depth" measurement for the depths and field sizes desired as well as by "in-air" or "in-water" calibration measurements for both constant SSD and isocentric techniques. Comparison of the above dose determination methods reveals the existence of both depth-dependent and field size-dependent dose discrepancies between the individual methods studied. For many teletherapy units, the source of these discrepancies lies in the inapplicability of the tabulated percent depth dose and tissue-air ratio values employed. Appropriate percent depth dose and tissue-air ratios can be obtained by performing a series of at-depth dose determinations for a variety of depths and field sizes. When the appropriate values are employed, excellent agreement is found between all of the dose determination methods investigated.Keywords