Detection of Hypoglycemia With the GlucoWatch Biographer
- 1 May 2001
- journal article
- clinical trial
- Published by American Diabetes Association in Diabetes Care
- Vol. 24 (5) , 881-885
- https://doi.org/10.2337/diacare.24.5.881
Abstract
OBJECTIVE—Hypoglycemia is a common acute complication of diabetes therapy. The GlucoWatch biographer provides frequent and automatic glucose measurements with an adjustable low-glucose alarm. We have analyzed the performance of the biographer low-glucose alarm relative to hypoglycemia as defined by blood glucose ≤3.9 mmol/l. RESEARCH DESIGN AND METHODS—The analysis was based on 1,091 biographer uses from four clinical trials, which generated 14,487 paired (biographer and blood glucose) readings. RESULTS—The results show that as the low-glucose alert level of the biographer is increased, the number of true positive alerts (alarm sounds and blood glucose ≤3.9 mmol/l) and false positive alerts (alarm sounds but blood glucose >3.9 mmol/l) increased. When analyzed as a function of varying low-glucose alert levels, the results show receiver operator characteristic curves consistent with a highly useful diagnostic tool. Setting the alert level from 1.1 to 1.7 mmol/l above the level of concern is likely to optimize the trade-off between true positives and false positives for each user. When the same blood glucose data are analyzed for typical monitoring practices (two or four measurements per day), the results show that fewer hypoglycemic events are detected than those detected with the biographer. CONCLUSIONS—The frequent and automatic nature of the biographer readings allows more effective detection of hypoglycemia than that achieved with current medical practice.Keywords
This publication has 16 references indexed in Scilit:
- Self-monitoring of blood glucose: language and financial barriers in a managed care population with diabetes.Diabetes Care, 2000
- The GlucoWatch®biographer: a frequent, automatic and noninvasive glucose monitorAnnals of Medicine, 2000
- Noninvasive Glucose MonitoringJAMA, 1999
- Application of the Mixtures of Experts algorithm for signal processing in a noninvasive glucose monitoring systemSensors and Actuators B: Chemical, 1999
- Accuracy of Glucose Reflectance Testing for Detecting Hypoglycemia in Term NewbornsClinical Pediatrics, 1999
- Design and Simulation of a Reverse Iontophoretic Glucose Monitoring DeviceJournal of the Electrochemical Society, 1998
- Self-Monitoring of Blood Glucose in Type I Diabetic Patients: Comparison With Continuous Microdialysis Measurements of Glucose in Subcutaneous Adipose Tissue During Ordinary Life ConditionsDiabetes Care, 1997
- Mathematical tools for demonstrating the clinical usefulness of biochemical markersScandinavian Journal of Clinical and Laboratory Investigation, 1997
- Measurement of glucose in diabetic subjects using noninvasive transdermal extractionNature Medicine, 1995
- Basic principles of ROC analysisSeminars in Nuclear Medicine, 1978