Wear and Surface Cracking in Early Retrieved Highly Cross-linked Polyethylene Acetabular Liners
Top Cited Papers
- 1 June 2004
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 86 (6) , 1271-1282
- https://doi.org/10.2106/00004623-200406000-00021
Abstract
Background: A higher degree of cross-linking has been shown to improve the tribological properties of ultra-high molecular weight polyethylene in laboratory studies; however, its effect on in vivo behavior has not been well established. We investigated in vivo wear mechanisms in retrieved highly cross-linked polyethylene acetabular liners in order to determine if early in vivo wear behavior is accurately predicted by hip-simulator studies. Methods: A total of twenty-four liners (twenty-one explanted and one unimplanted highly cross-linked liners and two explanted ethylene-oxide-sterilized non-cross-linked liners) were examined for this study. The average age of the patients was 59.9 years, and the average time in vivo was 10.1 months. Articular surface damage on the front and back sides of the liners was assessed with an optical scoring system. Surface quadrants were assigned a grade from 0 to 3 according to the observed wear mechanisms and the percentage of surface affected. The micromechanisms of liner damage were evaluated with use of scanning electron microscopy. Results: The average front and back-side explant damage scores were 11 (range, 2 to 26.5) and 6.7 (range, 3.7 to 13.3), respectively. There was consistent evidence of early surface deformation and cracking. All explants exhibited some form of surface change, including surface cracking, abrasion, pitting, or scratching. The original machining marks on the liner surface were observed to be either unaltered, drastically distorted, or absent. Conclusions: Highly cross-linked ultra-high molecular weight polyethylene acetabular liners that were retrieved at an average of ten months after implantation exhibited signs of surface damage that had not been predicted by in vitro hip-simulator studies. These devices had not failed clinically as a result of wear. The discrepancy between in vitro and in vivo wear surfaces may be due to variability in terms of in vivo lubrication and cyclic loading or may represent early surface damage mechanisms that are not well demonstrated by long-term simulator studies.Keywords
This publication has 10 references indexed in Scilit:
- A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical propertiesThe Journal of Arthroplasty, 2001
- Properties of crosslinked ultra-high-molecular-weight polyethyleneBiomaterials, 2000
- Effect of Sterilization Method and Other Modifications on the Wear Resistance of Acetabular Cups Made of Ultra-High Molecular Weight PolyethyleneJournal of Bone and Joint Surgery, 2000
- Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacementThe Journal of Arthroplasty, 1999
- Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethyleneJournal of Biomedical Materials Research, 1999
- Current Concepts Review - Wear in Total Hip and Knee Replacements*Journal of Bone and Joint Surgery, 1999
- A method of plant design evaluation featuring maintainability and reliabilityProceedings of the Institution of Mechanical Engineers, Part E: Journal of Process Mechanical Engineering, 1998
- Irradiation of chemically crosslinked ultrahigh molecular weight polyethyleneJournal of Polymer Science Part B: Polymer Physics, 1996
- Cold extrusion deformation of UHMWPE in total knee replacement prosthesesBiomaterials, 1995
- The appearances of wear on polyethylene—a comparison of in vivo and in vitro wear surfacesJournal of Biomedical Materials Research, 1978