Problems in the Use of Interference Filters for Spectrophotometric Determination of Total Ozone
- 1 August 1977
- journal article
- Published by American Meteorological Society in Journal of Applied Meteorology
- Vol. 16 (8) , 795-802
- https://doi.org/10.1175/1520-0450(1977)016<0795:pituoi>2.0.co;2
Abstract
An analysis of the use of ultraviolet narrow-band interference filters for total ozone determination is given with reference to the New Zealand filter spectrophotometer under the headings of filter monochromaticity, temperature dependence, orientation dependence, aging, and specification tolerances and non-uniformity. Quantitative details of each problem are given, together with the means used to overcome them in the New Zealand instrument. The tuning of the instrument's filter center wavelengths to a common set of values by tilting the filters is also described, along with a simple calibration method used to adjust and set these center wavelengths. Abstract An analysis of the use of ultraviolet narrow-band interference filters for total ozone determination is given with reference to the New Zealand filter spectrophotometer under the headings of filter monochromaticity, temperature dependence, orientation dependence, aging, and specification tolerances and non-uniformity. Quantitative details of each problem are given, together with the means used to overcome them in the New Zealand instrument. The tuning of the instrument's filter center wavelengths to a common set of values by tilting the filters is also described, along with a simple calibration method used to adjust and set these center wavelengths.Keywords
This publication has 0 references indexed in Scilit: