Assessing aortic valve area in aortic stenosis by continuity equation: a novel approach using real-time three-dimensional echocardiography
Open Access
- 9 February 2008
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 29 (20) , 2526-2535
- https://doi.org/10.1093/eurheartj/ehn022
Abstract
Two-dimensional echocardiographic (2DE) continuity-equation derived aortic valve area (AVA) in aortic stenosis (AS) relies on non-simultaneous measurement of left ventricular outflow tract (LVOT) velocity and geometric assumptions of LVOT area, which can amplify error, especially in upper septal hypertrophy (USH). We hypothesized that real-time three-dimensional echocardiography (RT3DE) can improve accuracy of AVA by directly measuring LVOT stroke volume (SV) in one window. RT3DE colour Doppler and 2DE were acquired in 68 AS patients (74 ± 12 yrs) prospectively. SV was derived from flow obtained from a sampling curve placed orthogonal to LVOT (Tomtec Imaging). Agreement between continuity-equation derived AVA by RT3DE (AVA3D-SV) and 2DE (AVA2D) and predictors of discrepancies were analysed. Validation of LVOT SV was performed by aortic flow probe in a sheep model with balloon inflation of septum to mimic USH. There was only modest correlation between AVA2D and AVA3D-SV (r = 0.71, difference 0.11 ± 0.23 cm2). The degree of USH was significantly associated with difference in AVA calculation (r = 0.4, P = 0.005). In experimentally distorted LVOT geometry in sheep, RT3DE correlated better with flow probe assessment (r = 0.96, P < 0.001) than 2DE (r = 0.71, P = 0.006). RT3DE colour Doppler-derived LVOT SV in the calculation of AVA by continuity equation is more accurate than 2D, including in situations such as USH, common in the elderly, which modify LVOT geometry.Keywords
This publication has 34 references indexed in Scilit:
- Real-time Three-dimensional Color Doppler Echocardiography Overcomes the Inaccuracies of Spectral Doppler for Stroke Volume CalculationJournal of the American Society of Echocardiography, 2006
- Absolute assessment of aortic valve stenosis by planimetry using cardiovascular magnetic resonance imaging: Comparison with transœsophageal echocardiography, transthoracic echocardiography, and cardiac catheterisationEuropean Journal of Radiology, 2006
- Diagnosis of aortic valvular stenosis by multislice cardiac computed tomographyEuropean Heart Journal, 2006
- Hemodynamic Assessment of Aortic StenosisJournal of the American College of Cardiology, 2005
- Flow-Dependent Changes in Doppler-Derived Aortic Valve Effective Orifice Area Are Real and Not Due to ArtifactJournal of the American College of Cardiology, 2005
- Real-time 3-Dimensional Doppler Echocardiography for the Assessment of Stroke Volume: An In Vivo Human Study Compared with Standard 2-Dimensional EchocardiographyJournal of the American Society of Echocardiography, 2005
- The use of live three-dimensional Doppler echocardiography in the measurement of cardiac outputJournal of the American College of Cardiology, 2005
- Usefulness of Live Three‐Dimensional Transthoracic Echocardiography in Aortic Valve Stenosis EvaluationThe American Journal of Geriatric Cardiology, 2004
- Three‐dimensional cross‐sectional velocity distribution in the ascending aorta in cardiac patientsClinical Physiology and Functional Imaging, 1996
- Noninvasive estimation of valve area in patients with aortic stenosis by Doppler ultrasound and two-dimensional echocardiography.Circulation, 1985