Some variables affecting reproducibility in capillary electrophoresis
- 1 January 1995
- journal article
- research article
- Published by Wiley in Electrophoresis
- Vol. 16 (1) , 2159-2163
- https://doi.org/10.1002/elps.11501601348
Abstract
Several variables that can affect reproducibility of both the peak height and the migration time in capillary electrophoresis (CE) were investigated here. A great part of the imprecision in CE for migration time is related to the interaction of the analyte with the capillary wall. Within a run, acidic compounds with long migration times have a higher relative standard deviation. In general, conditions that decreased the migration time, such as a short capillary length, tended to enhance the reproducibility of the migration time. Sample size also affected the reproducibility of peak height. Too small or too large an injection volume diminished the reproducibility. Sample matrix, such as high levels of proteins or salts in the sample, decreased the reproducibility. Furthermore, improvement in reproducibility can be achieved by using internal standards (especially for migration time), thorough washing of the capillary, or removing the excess proteins by acetonitrile deproteinization. Acetonitrile has an added effect of producing sample stacking. Depending on the migration time and analyte concentration, the peak height and area yielded different reproducibility results.Keywords
This publication has 10 references indexed in Scilit:
- Introduction of migration indices for identification: Chiral separation of some β‐blockers by using cyclodextrins in micellar electrokinetic capillary chromatographyElectrophoresis, 1994
- Quantitative applications and validation of the resolution of enantiomers by capillary electrophoresisElectrophoresis, 1994
- Sample matrix effects in capillary electrophoresis: II. Acetonitrile deproteinizationJournal of Chromatography A, 1993
- Sample matrix effects in capillary electrophoresis: I. Basic considerationsJournal of Chromatography A, 1993
- Impact of polycation adsorption on efficiency and electroosmotically driven transport in capillary electrophoresisAnalytical Chemistry, 1992
- Use of High Ionic Strength Buffers for the Separation of Proteins and Peptides with Capillary ElectrophoresisJournal of Liquid Chromatography, 1992
- Influence of operating parameters on reproducibility in capillary electrophoresisJournal of Chromatography A, 1991
- Internal standardization technique for capillary zone electrophoresisAnalytical Chemistry, 1991
- Capillary electrophoresis of proteins under alkaline conditionsJournal of Chromatography A, 1990
- Analysis of factors causing peak broadening in capillary zone electrophoresisJournal of Chromatography A, 1989