Evaluation of a smoke evacuator used for laser surgery
- 1 January 1989
- journal article
- research article
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 9 (3) , 276-281
- https://doi.org/10.1002/lsm.1900090311
Abstract
A preliminary study was conducted to determine the effectiveness of a smoke evacuation system used in laser surgery. A 30 W medical CO2 continuous wave (CW) laser was used to make cuts in a pork chop to simulate smoke production during laser surgery. A commercially available smoke evacuation system was used to control the smoke from the simulated surgery. The smoke concentration was measured at 6 in and at 3 and 4 ft from site of laser interaction. The nozzle of the smoke evacuator was located at distances of 2, 6, and 12 in from the surgical site to measure the relative effectiveness of the control. Complete control of smoke was achieved when the nozzle was located at 2 in, but significant amounts of smoke escaped when the nozzle was located at 6 and 12 in. Suggestions for the use of the smoke evacuation system and areas for further study are given.Keywords