Detection of aortic porcine valve dysfunction by maximum entropy spectral analysis.
- 1 July 1983
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 68 (1) , 42-49
- https://doi.org/10.1161/01.cir.68.1.42
Abstract
A high-resolution method of spectral analysis, of the class generally called "maximum entropy method," was used in a study of aortic porcine valve closing sounds in 37 patients (ages 19 to 76). Spectra from 27 normal xenografts, implanted from 2 weeks to 61 months previously, were characterized by a dominant frequency peak, F1, at 89 +/- 15 Hz (mean +/- SD), with a lower amplitude peak, F2, at 154 +/- 25 Hz. Eight of nine patients with aortic porcine valve dysfunction were proved surgically to have leaflet degeneration or infection and had either F1 (139 +/- 54 Hz) and/or F2 (195 +/- 74 Hz) significantly higher than normal (p less than .001). In two patients with paravalvar leak but no leaflet abnormality, F1 and F2 were in the normal range. Estimation of F1 and F2 was highly reproducible and was unaffected by duration of implant up to 5 years. Spectral analysis of aortic porcine valve closing sounds by the maximum entropy method may be useful for detection of intrinsic xenograft dysfunction.This publication has 12 references indexed in Scilit:
- Assessment of aortic regurgitation by means of pulsed Doppler ultrasoundUltrasound in Medicine & Biology, 1981
- Auscultation of the Normally Functioning Prosthetic ValveAnnals of Internal Medicine, 1981
- Spectrum of congestive heart failure late after aortic valve or mitral replacement: Differentiation of valvular versus myocardial cause by radionuclide ventriculogram-ejection fractionAmerican Heart Journal, 1981
- Pathological study of infective endocarditis on Hancock porcine bioprosthesesThe Journal of Thoracic and Cardiovascular Surgery, 1981
- Pathological substrates of thrombus formation after heart valve replacement with the Hancock bioprosthesisThe Journal of Thoracic and Cardiovascular Surgery, 1980
- Real-time sound spectroanalysis for diagnosis of malfunctioning prosthetic valvesThe Journal of Thoracic and Cardiovascular Surgery, 1980
- The Echocardiogram of the Porcine Aortic Bioprosthesis in the Aortic PositionChest, 1977