Performance of Seven Blood Glucose Testing Systems at High Altitude
- 1 October 1989
- journal article
- Published by SAGE Publications in The Diabetes Educator
- Vol. 15 (5) , 444-448
- https://doi.org/10.1177/014572178901500515
Abstract
Consumers and health care professionals expect blood glucose monitoring systems to consistently generate results that are close to actual blood glucose levels. Numerous environmental, physiologic, and operational factors can affect system performance, yielding results that are inaccurate or unpredictable. This study examined the effect of one factor—high altitude—on the performance of seven blood glucose monitoring systems. One of the systems overestimated blood glucose results; the other six systems underestimated blood glucose values (more than the expected variance). The findings of this study support previous reports of altered blood glucose monitoring system performance at high altitude. Diabetes educators can use this information when counseling consumers who reside or who plan to visit locations at high altitude.Keywords
This publication has 6 references indexed in Scilit:
- Identifying Sources of Error in Self-Monitoring of Blood GlucoseDiabetes Care, 1988
- Method for Estimating Confidence Levels for Measurements by Blood Glucose Monitoring SystemsDiabetes Care, 1988
- Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer IIDiabetes Care, 1988
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- Comparison of Various Methods for Rapid Glucose EstimationDiabetes Care, 1981